1,3-BUTADIENE: METHOD 1024, Issue 2, dated 15 August 1994 - Page 2 of 8



REAGENTS:

  1. Methylene chloride,[1] chromatographic quality with hydrocarbon (cyclohexene) preservative.
  2. 1,3-Butadiene,[1] 99.5%, in cylinder equipped for gas withdrawal, with needle valve.
  3. Helium, purified.
  4. Hydrogen, purified.
  5. Air, purified.
  6. Nitrogen, purified.
  7. Water, distilled.

EQUIPMENT:

  1. Sampler: Tandem charcoal tubes. Each tube is flame-sealed glass (8.5 cm long, 8-mm OD, 6-mm ID), has plastic caps for resealing, and contains activated coconut shell charcoal (such as SKC Lot 120) preceded by silylated glass wool and followed by a 3-mm urethane foam plug. The front tube holds 400 mg charcoal. The back tube holds 200 mg.
  2. Personal sampling pump, 0.01 to 0.5 L/min, with flexible connecting tubing./li>
  3. Refrigerant, bagged (e.g., Blue Ice or dry ice), and insulated shipping container.
  4. Gas chromatograph, flame ionization detector, integrator, and column (see APPENDIX A).
  5. Ice, wet.
  6. Vials, 5-mL, 2-mL, 1-mL, and other convenient sizes, with PTFE-lined septum caps.
  7. Pipettes, TD, 4-, 2-, and 1-mL.
  8. Syringes, gas-tight, 250-, 100-, 25-, and 10-µL.
  9. Beaker, 150-mL.
  10. Gas drying tube with serum cap to fit stem and 2-cm piece of plastic tubing to fit over serum cap.

SPECIAL PRECAUTIONS: 1,3-Butadiene is a potential carcinogen, teratogen, and reproductive hazard [2]. Methylene chloride is toxic, very volatile, and a suspect carcinogen [3]. Work should be performed in a well-ventilated fume hood.


SAMPLING:

  1. Calibrate each personal sampling pump with a representative sampler in line.
  2. Immediately before sampling, break ends of sampler tubes. Connect smaller tube to personal sampling pump with flexible tubing and to larger tube with a short piece of plastic tubing.
  3. Sample at an accurately known flow rate of 0.01 to 0.5 L/min for a sample size of 5 to 25 L.
  4. Separate the tubes, cap, and pack securely for shipment. Chill below –4 °C during shipment and storage.

SAMPLE PREPARATION:

  1. Add 4.0 mL methylene chloride to 5-mL vials and 2.0 mL to 2-mL vials. Loosely cap vials and thoroughly chill in ice.
  2. Place front sorbent sections in 5-mL vials and back sections in 2-mL vials. Discard glass wool and foam plugs. Immediately cap each vial.
  3. Remove from ice and allow to stand 30 min with occasional agitation.
  4. Transfer sample solution to appropriate vial and cap if using an autosampler. Thoroughly chill solution and vial before making transfer.

NIOSH Manual of Analytical Methods (NMAM), Fourth Edition

  1. 1.0 1.1 See SPECIAL PRECAUTIONS.