Page:NIOSH Manual of Analytical Methods - 1450.pdf/6
EST ERS 1: MET HO D 145 0, Issue 3, dated 15 M arch 200 3 - Page 6 o f 6 Table 2. Current Method Evaluation Overall Method
Storage Stability Study
Range (mg/m3 )
Accuracy
Breakthrough1 at 2 x OSHA PEL (# )
Bias (%)
S RT
Range Studied (µg/sample)
LOD (µg/sample)
Average DE
Measurement Precision (Ö r )
Levels (µg/sample)
Recovery (%)
n-amyl acetate
208-871
0.163
34.2 L
0.3
0.0831
14-440
0.9
0.96
0.0072
140
98
n-butyl acetate
352-1475
0.136
20.5 L
0.3
0.0691
15-440
0.9
0.96
0.0087
140
98
sec-butyl acetate
478-2005
0.116
16.5 L
-2.4
0.0539
14-440
0.9
0.97
0.0062
140
98
t-butyl acetate
424-1780
0.234
14.3 L
-8.6
0.0897
14-430
0.9
0.98
0.0052
140
98
2-ethoxyethyl acetate
262-1100
0.203
34.6 L
-9.6
0.0648
16-340
1
0.82
0.020
160
96
ethyl acrylate
50-210
0.162
>45 L
-7.1
0.0550
31-300
2
0.86
0.015
150
90
isoamyl acetate
208-874
0.195
32.3 L
-7.1
0.0750
14-430
0.9
0.97
0.0066
140
98
isobutyl acetate
306-1280
0.133
21.5 L
1.8
0.0656
14-440
0.9
0.97
0.0084
140
98
methyl isoamyl acetate
143-601
0.126
>45 L
-2.6
0.0590
14-280
0.9
0.81
0.015
140
95
n-propyl acetate
384-1610
0.162
17.9 L
6.9
0.0566
15-450
0.9
0.95
0.0091
140
98
Compound
1
Analytical Method
The values for these categories were taken from the original methods validation, see ref [1].
NIOSH Manual of Analytical Methods (NMAM), Fourth Edition