DICHLOROFLUOROMETHANE
CHCl2F
MW: 102.92
METHOD: 2516, Issue 2
CAS: 75-43-4
RTECS: PA8400000
EVALUATION: FULL
OSHA : 1000 ppm NIOSH: 10 ppm ACGIH: 10 ppm (1 ppm = 4.21 mg/m 3 @ NTP)
SYNONYMS:
2516
PROPERTIES:
Issue 1: 15 May 1985 Issue 2: 15 August 1994
gas; BP 8.9 °C; d 1.405 g/mL @ 9 °C; vapor density (air = 1) 3.82; not combustible
Freon 21; Halon 112; dichloromonofluoromethane
SAMPLING SAMPLER:
MEASUREMENT
SOLID SORBENT TUBES (2 coconut shell charcoal tubes, 400 mg and 200 mg)
TECHNIQUE:
GAS CHROMATOGRAPHY, FID
ANALYTE:
dichlorofluoromethane
FLOW RATE:
0.01 to 0.05 L/min
DESORPTION:
5 mL CS 2; stand 30 min
VOL-MIN: -MAX:
0.25 L @ 1000 ppm 3L
INJECTION VOLUME:
SHIPMENT:
separate front and backup tubes to avoid migration
TEMPERATURE-INJECTOR: -DETECTOR: -COLUMN:
SAMPLE STABILITY:
at least 7 days @ 25 °C
CARRIER GAS:
N 2, 30 mL/min
BLANKS:
2 to 10 field blanks per set
COLUMN:
stainless steel, 6 m x 3-mm, 10% FFAP on 100/120 mesh Chromosorb WHP
1730 to 7640 mg/m 3 [1] (3-L samples)
BIAS:
3.0%
ˆ ): 0.061 [1] OVERALL PRECISION (S rT ACCURACY:
125 °C 220 °C 55 °C
CALIBRATION:
dichlorofluoromethane solutions in CS 2 with internal standard
RANGE:
1 to 35 mg per sample
ESTIMATED LOD:
0.05 mg per sample
PRECISION (Sr):
0.024 @ 6 to 34 mg per sample [1]
ACCURACY RANGE STUDIED:
5 µL
± 13.3%
APPLICABILITY: The working range is 120 to 1500 ppm (500 to 6300 mg/m 3) for a 2-L air sample.
INTERFERENCES: None known.
OTHER METHODS: This revises Method S109 [2].
NIOSH Manual of Analytical Methods (NMAM), Fourth Edition, 8/15/94