Page:NIOSH Manual of Analytical Methods - 2518.pdf/1
HEXACHLORO-1,3-CYCLOPENTADIENE
C5Cl6
MW: 272.77
METHOD: 2518, Issue 2
CAS: 77-47-4
RTECS: GY1225000
EVALUATION: FULL
OSHA : no PEL NIOSH: 0.01 ppm; Group II Pesticide ACGIH: 0.01 ppm (1 ppm = 11.15 mg/m 3 @ NTP)
PROPERTIES:
2518
Issue 1: 15 May 1985 Issue 2: 15 August 1994
liquid; MP 10 °C; BP 239 °C; d 1.714 g/mL @ 15.5 °C; VP 13 Pa (0.1 mm Hg; 132 ppm) @ 25 °C
SYNONYMS: perchlorocyclopentadiene.
SAMPLING SAMPLER:
SOLID SORBENT TUBE (2 tubes, Porapak T, 75 mg and 25 mg)
FLOW RATE:
0.01 to 0.2 L/min
VOL-MIN: -MAX:
0.25 L @ 0.01 ppm 90 L
SHIPMENT:
separate tubes, seal, ship @ 25 °C; store @ 0 °C in the dark
SAMPLE STABILITY:
≥7 days @ 25 °C; ≥28 days @ 0 °C [1]
BLANKS:
2 to 10 field blanks per set
ACCURACY RANGE STUDIED:
13 to 873 µg/m 3 [1] (3-L samples)
BIAS:
- 0.3%
ˆ ): OVERALL PRECISION (S rT
0.082 [2]
ACCURACY:
± 16.4%
MEASUREMENT 63
TECHNIQUE:
GAS CHROMATOGRAPHY,
ANALYTE:
hexachloro-1,3-cyclopentadiene
DESORPTION:
1 mL hexane; ultrasonic bath 1 h
INJECTION VOLUME:
5 µL
TEMPERATURE-INJECTION: -DETECTOR: -COLUMN:
Ni ECD
150 °C 250 °C 135 °C
CARRIER GAS:
5% CH 4/95% Ar, 30 mL/min
COLUMN:
glass, 2 m x 4-mm ID; 3% OV-1 on 100/120 Gas Chrom Q
CALIBRATION:
hexachloro-1,3-cyclopentadiene in hexane
RANGE:
25 to 140 ng per sample
ESTIMATED LOD: 5 ng per sample PRECISION (Sr):
0.030 [1]
APPLICABILITY: The working range is 0.005 to 0.2 ppm (5 to 2400 mg/m 3) for a 5-L air sample. Area samples from a municipal wastewater treatment plant were analyzed by a variation of this method [2].
INTERFERENCES: None identified.
OTHER METHODS: This revises P&CAM 308 [3] which was evaluated with a reference method which involved sampling with impingers of hexane and gas chromatography with electron capture detection [1].
NIOSH Manual of Analytical Methods (NMAM), Fourth Edition, 8/15/94