Page:NIOSH Manual of Analytical Methods - 2558.pdf/1

This page has been proofread, but needs to be validated.
ACETOIN
2558

C4H8O2


MW: 88.11


CAS: 513-86-0


RTECS: EL8790000


METHOD: 2558, Issue 1


EVALUATION: PARTIAL


Issue 1: 15 March 2003


OSHA: None

NIOSH: None

ACGIH: None


PROPERTIES:


liquid, solid (dimer); d= 0.9972 g/mL @

20 °C; BP= 148 °C; FP = 50°C; MP=

15°C; Miscible in water and alcohol


SYNONYMS: 3-Hydroxy-2-butanone, 2,3-butanolone, acetyl methyl carbinol, dimethylketol


SAMPLING


MEASUREMENT


SAMPLER:
Anasorb CMS Solid Sorbent Tubes (150/75 mg)
FLOW RATE:
0.01 to 0.2 L/min
VOL-MIN:
1L
-MAX:
10 L
SHIPMENT:
Ship cold (5°C) and store in dark
SAMPLE
STABILITY:
30 days @ 5°C
BLANKS:
10% of field samples

ACCURACY


RANGE STUDIED:
Not Determined
BIAS:
Not Determined
OVERALL PRECISION(ŜrT):
Not Determined
ACCURACY:
Not Determined
TECHNIQUE:
GAS CHROMATOGRAPHY, FID
ANALYTE:
Acetoin
DESORPTION:
1 mL of acetone/methanol (95:5) on rotary mixer for 1.5 hours
INJECTION
VOLUME:
1µL
TEMPERATURE
-INJECTION:
225°C
-DETECTOR:
250°C
-COLUMN:
35°C (hold 3 min) to 200°C (6°C/min)
CARRIER GAS:
He (2.8 mL/min)
COLUMN:
Capillary, fused silica, 30 m x 0.32-mm
ID; 1-µm film, Stabilwax-DA or equivalent
CALIBRATION:
Standard solutions in acetoin in
acetone/methanol (95:5) solvent
RANGE:
3 to 378 µg
ESTIMATED LOD:
1.0 µg
PRECISION(S̄r):
0.008

APPLICABILITY: The working range for acetoin was 0.6 to 75.6 mg/m3 for a 5L air sample.


INTERFERENCES: Any compounds with similar retention times to acetoin.


OTHER METHODS: None determined.


NIOSH Manual of Analytical Methods (NMAM), Fourth Edition