ACETOIN
2558
|
C4H8O2
|
MW: 88.11
|
CAS: 513-86-0
|
RTECS: EL8790000 |
|
METHOD: 2558, Issue 1
|
EVALUATION: PARTIAL
|
Issue 1: 15 March 2003 |
|
OSHA: None NIOSH: None ACGIH: None
|
PROPERTIES:
|
liquid, solid (dimer); d= 0.9972 g/mL @ 20 °C; BP= 148 °C; FP = 50°C; MP= 15°C; Miscible in water and alcohol |
SYNONYMS: 3-Hydroxy-2-butanone, 2,3-butanolone, acetyl methyl carbinol, dimethylketol
|
SAMPLING
|
MEASUREMENT |
|
SAMPLER: Anasorb CMS Solid Sorbent Tubes (150/75 mg) FLOW RATE: 0.01 to 0.2 L/min VOL-MIN: 1L -MAX: 10 L SHIPMENT: Ship cold (5°C) and store in dark SAMPLE STABILITY: 30 days @ 5°C BLANKS: 10% of field samples ACCURACY RANGE STUDIED: Not Determined BIAS: Not Determined OVERALL PRECISION(ŜrT): Not Determined ACCURACY: Not Determined |
TECHNIQUE: GAS CHROMATOGRAPHY, FID ANALYTE: Acetoin DESORPTION: 1 mL of acetone/methanol (95:5) on
rotary mixer for 1.5 hours INJECTION VOLUME: 1µL TEMPERATURE -INJECTION: 225°C -DETECTOR: 250°C -COLUMN: 35°C (hold 3 min) to 200°C (6°C/min) CARRIER GAS: He (2.8 mL/min) COLUMN: Capillary, fused silica, 30 m x 0.32-mm ID; 1-µm film, Stabilwax-DA or equivalent CALIBRATION: Standard solutions in acetoin in acetone/methanol (95:5) solvent RANGE: 3 to 378 µg ESTIMATED LOD: 1.0 µg PRECISION(S̄r): 0.008 |
APPLICABILITY: The working range for acetoin was 0.6 to 75.6 mg/m3 for a 5L air sample.
INTERFERENCES: Any compounds with similar retention times to acetoin.
OTHER METHODS: None determined.
NIOSH Manual of Analytical Methods (NMAM), Fourth Edition