Page:NIOSH Manual of Analytical Methods - 3502.pdf/1
PHENOL
C6H5OH
MW: 94.11
METHOD: 3502, Issue 2
CAS: 108-95-2
RTECS: SJ3325000
EVALUATION: FULL
Issue 1: 15 February 1984 Issue 2: 15 August 1993
SYNONYMS: carbolic acid.
SAMPLING SAMPLER:
BUBBLER (0.1 N sodium hydroxide)
MEASUREMENT TECHNIQUE:
GAS CHROMATOGRAPHY, FID
ANALYTE:
phenol
pH ADJUSTMENT:
0.1 mL conc. sulfuric acid; pH < 4
INJECTION VOLUME:
5 µL
FLOW RATE: 0.2 to 1 L/min VOL-MIN: -MAX:
26 L 240 L
SHIPMENT:
hand delivery or special bubbler shipping cases
SAMPLE STABILITY:
at least 5 days @ 25 C [1]
BLANKS:
2 to 10 field blanks per set
TEMPERATURE-INJECTION: -DETECTOR: -COLUMN:
215 C 225 C 200 C
CARRIER GAS:
N 2 or He, 50 mL/min
COLUMN:
stainless steel, 1.2 m x 6-mm OD, 35/60 mesh Tenax or equivalent
CALIBRATION:
phenol in 0.1 N sodium hydroxide
RANGE:
0.5 to 6 mg [2]
ACCURACY RANGE STUDIED: 9.5 to 38 mg/m 3 [1] (100-L samples) BIAS:
-5.87% [1]
ESTIMATED LOD: 10 µg per sample [3]
ˆ rT): 0.068 [1] OVERALL PRECISION (S ACCURACY:
+ 16.5%
PRECISION (Sr):
0.044 [1]
APPLICABILITY: The working range is 5 to 60 mg/m 3 for a 100-L air sample [4]. This method may be used for STEL measurements (15-L samples).
INTERFERENCES: None identified. Ethanol does not interfere.
OTHER METHODS: This is Method S330 in a revised format [2]; the method also appears in the phenol criteria document [4].
NIOSH Manual of Analytical Methods (NMAM), Fourth Edition, 8/15/93