TETRAMETHYL THIOUREA
(CH3)2NC(=S)N(CH3)2
MW: 132.23
METHOD: 3505, Issue 2
3505
CAS: 2782-91-4
EVALUATION: PARTIAL
OSHA : no standard NIOSH: no recommended exposure limit ACGIH: no standard
RTECS: YU2750000 Issue 1: 15 May 1985 Issue 2: 15 August 1994
PROPERTIES: solid; MP 78 °C; BP 245 °C
SYNONYMS: 1,1,3,3-tetramethyl-2-thiourea; TMTU
SAMPLING SAMPLER:
IMPINGER (water)
MEASUREMENT TECHNIQUE:
VISIBLE ABSORPTION SPECTROPHOTOMETRY
ANALYTE:
tetramethyl thiourea-pentacyanoamineferrate coordination complex
WAVELENGTH:
590 nm
CALIBRATION:
solutions of tetramethyl thioureapentacyandamine ferrate coordination complex in water
RANGE:
15 to 150 µg per sample
FLOW RATE: 0.2 to 1 L/min VOL-MIN: -MAX:
50 L @ 0.3 mg/m 3 250 L
SHIPMENT:
ship impinger contents in vial
SAMPLE STABILITY:
unknown
BLANKS:
2 to 10 field blanks per set
BULK SAMPLE:
required; high-volume air or settled dust
ESTIMATED LOD: 3 µg per sample [1] PRECISION (Sr):
0.02 @ 15 to 150 µg per sample [1,2]
ACCURACY RANGE STUDIED:
generated atmospheres not studied
BIAS:
unknown
ˆ ): not determined OVERALL PRECISION (S rT ACCURACY:
not determined
APPLICABILITY: The working range is 0.15 to 10 mg/m 3 for a 100-L air sample. Tetramethyl thiourea is used in the manufacture of adhesives and polychloroprene rubber. The method was also applied to ethylene thiourea [1].
INTERFERENCES: Compounds containing a thione (C=S) group will complex with the pentacyanoamineferrate reagent and may interfere with the analyte absorbance band at 590 nm. Other compounds that absorb at 590 nm may interfere.
OTHER METHODS: This method was originally designated P&CAM 282 [1], which it replaces.
NIOSH Manual of Analytical Methods (NMAM), Fourth Edition, 8/15/94