Page:NIOSH Manual of Analytical Methods - 5009.pdf/1
BENZOYL PEROXIDE
(C6H5CO)2O2
MW: 242.22
METHOD: 5009, Issue 2
CAS: 94-36-0
5009
RTECS: DM8575000
EVALUATION: FULL
OSHA : 5 mg/m 3 NIOSH: 5 mg/m 3 ACGIH: 5 mg/m 3
PROPERTIES:
Issue 1: 15 February 1984 Issue 2: 15 August 1994
solid; MP 103 °C; d 1.334 g/mL @ 25 °C VP not significant; autoignition temperature 80 °C
SYNONYMS: dibenzoyl peroxide, benzoyl superoxide
SAMPLING
MEASUREMENT
SAMPLER:
FILTER (0.8-µm cellulose ester membrane)
TECHNIQUE:
HPLC, UV DETECTION
FLOW RATE:
1 to 3 L/min
ANALYTE:
benzoyl peroxide
VOL-MIN: -MAX:
40 L @ 5 mg/m 3 400 L
EXTRACTION:
ethyl ether, 10 mL
INJECTION VOLUME: 20 µL
SHIPMENT:
refrigerated
PRESSURE-COLUMN: 9000 kPa (1300 psi)
SAMPLE STABILITY:
9% loss from filter after 1 week @ 25 °C
MOBILE PHASE:
70/30 methanol/water; 1.6 mL/min
DETECTOR:
UV photometer @ 254 nm
BLANKS:
2 to 10 field blanks per set COLUMN:
250 mm x 3 mm-ID stainless steel; Spherisorb ODS (spherical silica particles with 5% bonded coating of octadecyl groups) or equivalent
CALIBRATION:
benzoyl peroxide in ethyl ether
ACCURACY RANGE STUDIED:
3 to 19 mg/m 3 [1] (90-L samples)
BIAS:
- 0.52%
ˆ ): OVERALL PRECISION (S rT
RANGE:
0.2 to 1.7 mg per sample [1]
ESTIMATED LOD:
0.01 mg per sample [2]
PRECISION (Sr):
0.024 [1]
0.060 [1]
ACCURACY:
± 11.82%
APPLICABILITY: The working range is 2 to 19 mg/m 3 for a 90-L air sample.
INTERFERENCES: None identified.
OTHER METHODS: This is Method S253 [3] in a revised format. A non-specific gravimetric method and a non-specific colorimetric method appear in the criteria document [4].
NIOSH Manual of Analytical Methods (NMAM), Fourth Edition, 8/15/94