2,4,7-TRINITROFLUOREN-9-ONE
C13 H5N3O7
MW: 315.20
METHOD: 5018, Issue 2
CAS: 129-79-3
5018
RTECS: LL9100000
EVALUATION: PARTIAL
OSHA : no PEL NIOSH: no REL ACGIH: no TLV
Issue 1: 15 May 1985 Issue 2: 15 August 1994
PROPERTIES: solid yellow needles; MP 176 °C
SYNONYMS: TNF
SAMPLING SAMPLER:
FILTER (0.5-µm, PTFE membrane)
FLOW RATE:
1 to 3 L/min
VOL-MIN: -MAX:
100 L @ 2 µg/m 3 500 L
SHIPMENT:
routine
MEASUREMENT TECHNIQUE:
HPLC, UV DETECTION
ANALYTE:
2,4,7-trinitrofluoren-9-one
PREPARATION:
extract with 2 mL toluene; centrifuge
INJECTION VOLUME:
100 µL
MOBILE PHASE: 20% isooctane/80% methylene chloride; 2 mL/min; ambient temperature
SAMPLE STABILITY:
at least 2 weeks @ 25 °C
BLANKS:
2 to 10 field blanks per set
COLUMN:
Waters Radial PAK B (packed with 10µm silica) with Radial Compression Module, or equivalent
DETECTOR WAVELENGTH:
280 nm
CALIBRATION:
standard solutions of TNF in toluene
RANGE:
0.2 to 4 µg per sample
ACCURACY RANGE STUDIED:
not studied
BIAS:
none identified ESTIMATED LOD: 0.04 µg per sample [1]
ˆ ): not evaluated OVERALL PRECISION (S rT PRECISION (Sr): ACCURACY:
0.056 [1]
not determined
APPLICABILITY: The working range is 0.4 to 100 µg/m 3 for a 500-L air sample. determination of TNF in spent toner from business copy machines.
This method was developed for the
INTERFERENCES: None known.
OTHER METHODS: This revises P&CAM 348 [1].