Page:NIOSH Manual of Analytical Methods - 5036.pdf/1
TRIMELLITIC ANHYDRIDE
C9H4O 5
MW: 192.13
METHOD: 5036, Issue 1
CAS: 552-30-7
5036
RTECS: DC2050000
EVALUATION: PARTIAL
OSHA : No PEL NIOSH: 0.005 ppm/10 h ACGIH: 0.005 ppm (1 ppm = 7.86 mg/m 3 @ NTP)
PROPERTIES:
Issue 1: 15 August 1994
solid; MP 161 to 163 °C; VP 5.3 x 10 -4 Pa (4 x 10 -6 mm Hg) @ 20 °C
SYNONYMS: anhydrotrimellitic acid; 1,3-dioxo-phthalancarboxylic acid; 1,3-dihydro-1,3-dioxo-5-isobenzofurancarboxylic acid; 1,2,4-benzenetricarboxylic acid anhydride; 1,2,4-benzenetricarboxylic acid cyclic-1,2-anhydride. SAMPLING
MEASUREMENT
SAMPLER:
FILTER (0.8-µm PVC-copolymer)
TECHNIQUE:
GAS CHROMATOGRAPHY, FID
FLOW RATE:
1.5 to 2.0 L/min
EXTRACTION:
methanol, 3 mL
VOL-MIN: -MAX:
400 L not specified
DERIVATIZATION:
SHIPMENT:
routine
SAMPLE STABILITY:
unknown
BLANKS:
2 to 10 field blanks per set
INJECTION VOLUME:
boron trifluoride/methanol
5 µL
TEMPERATURE-INJECTION: -DETECTOR: -COLUMN:
315 325 250 310
°C °C °C, 2 min; 16 °C/min; °C, 4 min
COLUMN:
stainless steel, tenax 60/80 mesh, (9 ft x 2-mm)
CARRIER GAS:
Helium, 35 mL/min
CALIBRATION:
standard solutions of trimethyl trimellitate (derivative) in a mixture of pyridine, diethyl ether, and pentane
RANGE:
19 to 96 µg per sample
ACCURACY RANGE STUDIED:
0.048 to 0.24 mg/m 3 [1] (400-L samples)
BIAS:
not significant
ˆ ): not determined OVERALL PRECISION (S rT
ESTIMATED LOD: 2 µg per sample
ACCURACY:
PRECISION (Sr):
not determined
0.087 @ 19.2 to 96 µg per sample
APPLICABILITY: The working range is 0.048 to 0.24 mg/m 3 for a 400-L air sample. The method was used for analysis of bulk samples and air (field samples) [2,3]. This method will not differentiate between trimellitic anhydride and trimellitic acid [1].
INTERFERENCES: Any compound with the same retention time is an interference. The interference may be eliminated by changing the temperature program or carrier gas flow rate.
OTHER METHODS: This revises Method P&CAM 322 [1], which was based on a procedure by Biondi and Cagnosso [4].
NIOSH Manual of Analytical Methods (NMAM), Fourth Edition, 8/15/94