Page:NIOSH Manual of Analytical Methods - 5502LIND.pdf/1
LINDANE See ALDRIN, Method 5502, for Procedure
C6H6Cl6
MW: 290.82
CAS: 58-89-9
METHOD: 5502, Issue 2
RTECS: GV4900000
EVALUATION: FULL
OSHA : 0.5 mg/m 3 (skin) NIOSH: 0.5 mg/m 3 (skin) ACGIH: 0.5 mg/m 3 (skin)
5502
Issue 1: 15 February 1984 Issue 2: 15 August 1994
PROPERTIES: solid; MP 112.5 °C; VP 0.0013 Pa (9.4 x 10 -6 mm Hg; 0.15 mg/m 3) @ 20 °C
SYNONYMS: gamma-hexachlorocyclohexane; benzenehexachloride; gamma-BHC.
SAMPLING
MEASUREMENT
SAMPLER:
FILTER AND BUBBLER (glass fiber + 15 mL isooctane)
TECHNIQUE: GAS CHROMATOGRAPHY, ELECTROLYTIC CONDUCTIVITY DETECTOR
FLOW RATE:
0.2 to 1 L/min
ANALYTE:
VOL-MIN: -MAX:
18 L @ 0.25 mg/m 3 240 L
FILTER EXTRACTION:
isooctane
SHIPMENT:
transfer bubbler solutions and filters in scintillation vials; pack carefully
INJECTION VOLUME:
15 µL
SAMPLE STABILITY:
at least 1 week @ 25 °C
BLANKS:
2 to 10 field blanks per set
LINDANE
TEMPERATURE-FURNACE: -TRANSFER: -VENT: -COLUMN: GASES-H2 (furnace): -N2 (carrier):
ACCURACY RANGE STUDIED:
0.3 to 1.7 mg/m 3 [1];
BIAS:
- 0.44%
COLUMN:
750 225 205 160
to 770 °C °C to 260 °C to 190 °C
150 to 160 mL/min 140 mL/min
glass, 1.2 m x 3-mm OD; 5% SE-30 on 80/100 mesh acid-washed DMCS Chromosorb W or equivalent
CALIBRATION:
solution of analyte in isooctane
ˆ rT): 0.086 [1] OVERALL PRECISION (S
RANGE:
5 to 135 µg per sample
ACCURACY:
ESTIMATED LOD: 3 µg per sample
± 16.9%
PRECISION (Sr):
0.013 [1]
APPLICABILITY: The working range is 0.05 to 1.5 mg/m 3 for a 90-L air sample. Evaporation of isooctane from the bubbler necessitates refilling the bubbler frequently.
INTERFERENCES: None identified.
OTHER METHODS: This method combines and replaces S275 [2] and S290 [2]. Lindane has also been sampled with a filter-solid sorbent train [3].
NIOSH Manual of Analytical Methods (NMAM), Fourth Edition, 8/15/94