Page:NIOSH Manual of Analytical Methods - 5519.pdf/1
ENDRIN
C12 H8OCl6
MW: 380.93
METHOD: 5519, Issue 2
5519
CAS: 72-20-8
EVALUATION:
OSHA : 0.1 mg/m 3 (skin) NIOSH: 0.1 mg/m 3 (skin); Group I Pesticide ACGIH: 0.1 mg/m 3 (skin)
RTECS: IO1575000
FULL
PROPERTIES:
Issue 1: 15 May 1989 Issue 2: 15 August 1994
crystals; MP 200 °C; decomposes @ 245 °C; VP 0.27 x 10 -7 kPa (2 x 10 -7 mm Hg) @ 25 °C
SYNONYMS: Mendrin; Nendrin; Hexadrin
SAMPLING SAMPLER:
MEASUREMENT
FILTER + SOLID SORBENT TUBE (0.8 µm cellulose ester membrane + Chromosorb 102, 100/50 mg)
TECHNIQUE:
GAS CHROMATOGRAPHY,
ANALYTE:
endrin
FLOW RATE:
0.5 to 1 L/min
EXTRACTION:
5 mL toluene, stand 15 min
VOL-MIN: -MAX:
12 L @ 0.1 mg/m 3 400 L
INJECTION VOLUME:
SHIPMENT:
routine
SAMPLE STABILITY:
at least 1 week @ 25 °C [1]
BLANKS:
2 to 10 field blanks per set
63
Ni ECD
5 µL
TEMPERATURE-INJECTOR: -DETECTOR: -COLUMN:
175 °C 280 °C 160 °C
CARRIER GAS:
95% argon/5% methane @ 60 mL/min
COLUMN:
2 m x 4-mm ID glass, packed with 3% OV-1 on 100/120 Chromosorb Q
CALIBRATION:
standard solutions of endrin in toluene
RANGE:
1.2 to 36 µg per sample [2]
ACCURACY RANGE STUDIED:
0.06 to 0.31 mg/m 3 [1,2] (120-L samples)
BIAS:
- 0.7%
ˆ ): 0.071 [2] OVERALL PRECISION (S rT ACCURACY:
ESTIMATED LOD: 0.02 µg per sample [1] PRECISION (Sr):
0.016 @ 1.2 to 24.5 µg per sample [1]
± 13.9%
APPLICABILITY: The working range is 0.01 to 0.33 mg/m 3 for a 120-L air sample. Smaller concentrations can be determined if desorption efficiency is satisfactory. The use of a capillary column, e.g., DB-1, may improve resolution and sensitiv ity.
INTERFERENCES: None identified.
OTHER METHODS: This is a modification of the method of Hill and Arnold [3] and it replaces NIOSH method S284 [2].
NIOSH Manual of Analytical Methods (NMAM), Fourth Edition, 8/15/94