Page:NIOSH Manual of Analytical Methods - 5603.pdf/5
ALACHLOR in AIR: METHOD 5603, Issue 1, dated 15 January 1998 - Page 5 of 5
REFERENCES: [1] Sammons DL, Kennedy ER [1997]. Backup data for alachlor (unpublished report). [2] Aga DS, Thurman EM, Pomes ML. Determination of alachlor and its sulfonic acid metabolite in water by solid-phase extraction and enzyme-linked immunosorbent assay.Analytical Chemistry. 66:14951499 (1994). [3] NIOSH [1998]. Chlorinated organonitrogen and acid herbicides in air: Method 5602. In: Eller PM, Cassinelli ME, eds. NIOSH manual of analytical methods, th4 ed., 2nd supplement. Cincinnati, OH: National Institute for Occupational Safety and Health, DHHS (NIOSH) Publication No. 98-119. [4] EnviroGard Alachlor Plate Kit. P 30128, Rev A 3/14/95. Millipore. [5] Sanderson WT, Biagini R, Henningsen G, Ringenburg V, MacKenzie B [1995]. Exposure of commercial pesticide applicators to the herbicide alachlor. Am Ind Hyg Assoc 56:890-897. J
METHOD WRITTEN BY: Deborah Sammons and Eugene Kennedy, Ph.D., NIOSH/DPSE
TABLE 1. METHOD SUMMARY Method Evaluation Data Test Average (µg/sample)
Independent Average (µg/sample)
Precisionb Bias Calculated (sampling and [(test/indep) -1] Accuracy (± %) analysis)
0.01x
0.259
0.24
0.109
0.080
27.7
0.1x
2.17
2.4
0.170
-0.094
35.3
0.5x
12.7
12.0
0.081
0.059
23.0
1x
23.9
24.0
0.063
-0.003
13.2
2x
48.9
48.0
0.090
0.019
18.4
5x
128.4
120
0.122
0.070
29.1
Arbitrary Exposure Limit (1 mg/m3)a
a
b
Theoretical Exposure Concentrations were based on the NIOSH RELs of similarly classified pesticides. Precision is not statistically poolable.
NIOSH Manual of Analytical Methods (NMAM), Fourth Edition