Page:NIOSH Manual of Analytical Methods - 6007.pdf/1
NICKEL CARBONYL
Ni(CO)4
MW: 170.73
METHOD: 6007, Issue 2
CAS: 13463-39-3
6007
RTECS: QR6300000
EVALUATION: PARTIAL
OSHA : 0.001 ppm (as Ni) NIOSH: 0.001 ppm; carcinogen ACGIH: 0.05 ppm (as Ni) (1 ppm = 6.98 mg/m 3 @ NTP)
Issue 1: 15 August 1987 Issue 2: 15 August 1994
PROPERTIES: liquid; d 1.318 g/mL @ 20 °C; BP 43 °C; MP -25 °C; VP 43 kPa (321 mm Hg; 42% v/v) @ 20 °C; lower explosive limit 2% v/v in air
SYNONYMS: nickel tetracarbonyl
SAMPLING SAMPLER:
PREFILTER + SOLID SORBENT TUBE (low-Ni charcoal, 120 mg/60 mg; 0.8-µm cellulose ester membrane)
MEASUREMENT TECHNIQUE:
ATOMIC ABSORPTION, GRAPHITE FURNACE
ANALYTE:
nickel
DESORPTION:
1 mL 3% HNO 3; 30 min ultrasonic bath
INJECTION VOLUME:
20 µL
FLOW RATE: 0.05 to 0.2 L/min VOL-MIN: -MAX:
7 L @ 0.001 ppm 80 L
SHIPMENT:
routine
SAMPLE STABILITY:
BLANKS:
GRAPHITE FURNACE: 95% recovered after 17 days @ room temperature [1]
110 °C dry 30 sec; 800 °C char 15 sec; ≥ 2700 °C atomize 10 sec
WAVELENGTH:
232 nm, with background correction
CALIBRATION:
standard solution of Ni 2+ in 3% (w/v) HNO 3
RANGE:
0.05 to 0.6 µg Ni per sample [1]
2 to 10 field blanks per set
ACCURACY RANGE STUDIED:
0.0007 to 0.017 ppm [1] (3- to 40-L air samples)
BIAS:
- 7% [1]
ESTIMATED LOD: 0.01 µg Ni per sample [1] PRECISION (Sr):
0.028 @ 0.08 to 0.5 µg Ni per sample
ˆ ): 0.099 [1] OVERALL PRECISION (S rT ACCURACY:
± 26.4%
APPLICABILITY: The working range is 0.00036 to 0.007 ppm (0.0025 to 0.05 mg/m 3) nickel carbonyl for a 20-L air sample.
INTERFERENCES: Particulate nickel compounds will give a positive interference unless a prefilter is used during sampling.
OTHER METHODS: This revises P&CAM 344 [2]. A colorimetric method has also been recommended [3].
NIOSH Manual of Analytical Methods (NMAM), Fourth Edition, 8/15/94