Page:NIOSH Manual of Analytical Methods - 7304.pdf/11
ELEMENTS by ICP (Microwave Digestion): METHOD 7304, Issue 1, dated 25 June 2014 - Page 11 of 11
Table 4. OVERALL PRECISION AND ACCURACY DATA [13] Range Studied (µg/sample)
Precision Accuracy (%) SrT
Lowest Level(2)
0.0163
0.0155
<5
10.05
0.0334
0.0407
8.2
15
Instrument
From
To
From
To
Tellurium
P-E Optima
10.05
3000
-0.0043
-0.0282
Thallium
Fisons
5.025
1500
-0.0081
-0.0362
Thallium
P-E Optima
5.025
1500
Fisons P-E Optima
25.12
7500
-0.0505 -0.0688 -0.0048 0.0250 9 Poor and variable recoveries across study range.
25.12
7500
Poor and variable recoveries across study range.
Tin Tin Titanium
Bias
Range of Bias
Element
(1)
5.025
Fisons
1.005
300
-0.0827
-0.1834
0.0342
0.0269
12.3
3
Titanium
P-E Optima
1.005
300
-0.1072
-0.1732
-0.0387
0.0321
15.3
1.005
Vanadium
Fisons
0.5038
150.4
0.0704
0.0438
0.1114
0.0195
10.5
0.5038
Vanadium
P-E Optima
0.5038
150.4
-0.0063
-0.0217
0.0099
0.0198
<5
0.5038
Yttrium
Fisons
0.2519
75.2
0.0598
0.0466
0.0795
0.0164
8.9
0.2519
Zinc
Fisons
1.005
300
0.1452
0.0630
0.2976
0.0340
22
1.005
Zinc
Fisons
1.005
300
0.1190
0.0630
0.1683
0.0356
18.7
3
Zinc
P-E Optima
1.005
300
-0.0502
-0.0655
-0.0388
0.0295
9.6
3
Zirconium
Fisons
0.5025
150
0.0164
-0.0096
0.0350
0.0175
<5
0.5025
(1) Values reported were obtained with a Fisons ARL Accuris ICP-AES or a Perkin Elmer Optima 3000 DV ICP-AES. (2) Lowest level in range studied at which recoveries were between 81 and 121% recovery and relative standard deviation (Sr) less than 0.1100 on 5 or 6 replicates. Performance may vary with instrument and should be independently verified.
NIOSH Manual of Analytical Methods (NMAM), Fifth Edition