LEAD SULFIDE
PbS
MW: 239.25
METHOD: 7505, Issue 2
CAS: 1314-87-0
7505
RTECS: OG4550000
EVALUATION: PARTIAL
OSHA : 0.05 mg (Pb)/m 3 NIOSH: <0.1 mg (Pb)/m 3; blood Pb ≤ 60 µg/100 g ACGIH: 0.15 mg (Pb)/m 3
Issue 1: 15 February 1984 Issue 2: 15 August 1994
PROPERTIES: solid; d 7.5 g/mL; MP 1114 °C
SYNONYMS: galena (mineral)
SAMPLING SAMPLER:
CYCLONE + FILTER (10-mm nylon or Higgins-Dewell (HD) cyclone + 5-µm PVC membrane)
FLOW RATE: HD cyclone: 2.2 L/min nylon cyclone: 1.7 L/min
MEASUREMENT TECHNIQUE:
X-RAY POWDER DIFFRACTION
ANALYTE:
lead sulfide
DISSOLVE:
filter in tetrahydrofuran
SUSPEND:
dust in tetrahydrofuran
VOL-MIN: -MAX:
600 L @ 0.05 mg (Pb)/m 3 1000 L
REDEPOSIT:
0.45-µm Ag membrane filter
SHIPMENT:
routine
XRD:
Cu target X-ray tube Optimize for intensity; 1° receiving slit Graphite monochromator; scintillation detector Slow step scan, 0.02°/10 sec Integrated intensity with background subtraction
CALIBRATION:
suspensions of PbS in 2-propanol
RANGE
30 to 2000 µg per sample
SAMPLE STABILITY:
4% degradation in 45 days; protect samples from light
BLANKS:
2 to 10 field blanks per set
BULK SAMPLE:
required; area respirable or settled dust
ACCURACY RANGE STUDIED: BIAS:
12 to 37 µg/m 3 [1] not significant when samples and standards have equivalent particle size
ESTIMATED LOD: 5 µg per sample PRECISION (Sr):
0.081 (40 to 260 µg per sample [1]
ˆ ): 0.103 [1] OVERALL PRECISION (S rT ACCURACY:
ca. ± 22%
APPLICABILITY: The working range is 0.06 to 4 mg/m 3 for a 500-L air sample. The method is specific for lead sulfide among other mine dusts.
INTERFERENCES: Lead oxide (PbO, Pb 3O 4), lead sulfate (anglesite), copper iron sulfide (chalcopyrite); also see APPENDIX.
OTHER METHODS: This method was originally designated P&CAM 350, which it replaces [2].
NIOSH Manual of Analytical Methods (NMAM), Fourth Edition, 8/15/94