Page:NIOSH Manual of Analytical Methods - 9102.pdf/5
ELEM ENT S O N W IPES: M ETH OD 9102, Issue 1, dated 1 5 Ma rch 2003 - Page 5 of 5
TABLE 2. WAVELENGTH, LIMITS OF DETECTION (LOD), RANGE & PRECISION W avelen gth (nm )
Instrumental LOD (µg/wipe)
Range (µg/wipe)
Precision* (S r)
Silver (Ag)
328
0.035
0.11 8-25 .0
0.0701
Arsenic (As)
189
0.078
0.261-105.
0.0549
Barium (Ba)
455
0.010
0.03 33-1 0.0
0.0370
Beryllium (Be)
313
0.010
0.03 33-1 0.0
0.0267
Cadm ium (Cd)
226
0.010
0.03 33-4 1.7
0.0273
Cobalt (Co)
228
0.010
0.03 33-1 0.0
0.0301
Chrom ium (C r)
267
0.010
0.03 33-1 0.0
0.0287
Copper (Cu)
324
0.042
0.141-114.
0.0317
Iron (Fe)
259
0.074
0.24 6-50 .0
0.0516
Lanthanum (La)
408
0.018
0.05 93-2 5.0
0.0279
Manganese (Mn)
257
0.010
0.0333-638.
0.0333
Molybdemum (Mo)
202
0.010
0.03 33-1 0.0
0.0441
Nick el (N i)
231
0.010
0.03 40-2 0.6
0.0629
Phosphorus (P)
178
0.043
0.142-860.
0.128
Lead (Pb)
168
0.042
0.141-1160
0.0267
Selenium (Se)
196
0.10
0.33 8-75 .0
0.0364
Strontium (S r)
407
0.010
0.03 33-1 0.0
0.0368
Tellurium (Te)
214
0.047
0.15 6-50 .0
0.0359
Thallium (Tl)
190
0.041
0.13 6-50 .0
0.0256
Vanadium (V)
292
0.010
0.03 33-2 5.0
0.0284
Yttrium (Y)
371
0.010
0.03 33-1 0.0
0.0280
Zinc (Zn)
213
0.77
2.56-350.
0.101
Zirconium (Zr)
339
0.026
0.08 77-1 0.0
0.344
Elem ent (Sym bo l)
- Values were obtained with a Spectro ICP-AES; LOD may vary with instrument and should be
independently verified.
NIOSH Manual of Analytical Methods (NMAM), Fourth Edition