Page:NIOSH Manual of Analytical Methods - 9106.pdf/18
METHAMPHETAMINE . . .on Wipes by Liquid-Liquid Extraction: METHOD 9106, Issue 1, dated 17 October 2011 - Page 18 of 31
Table 8b. Precision and accuracy in SIM mode.(1) Compound (D)-Amphetamine
Internal Range(3) µg/ Standard(2) sample Accuracy(4)
Overall(4) Precision
Bias Average
Range
D11-Amp
0.1-30
14.3
0.0412
-0.0750
-0.1153 - -0.0351
D14-Met
0.1-30
10.1
0.0508
-0.0074
-0.0500 - +0.0389
NMPhen
0.1-30
13.3
0.0439
-0.0606
-0.1117 - -0.0318
D11-Amp
0.1-30
21.3
0.0578
-0.1182
-0.1949 - -0.0697
D14-Met
0.1-30
14.4
0.0534
-0.0558
-0.1061 - -0.0170
NMPhen
0.3-30
19.8
0.0387
-0.1338
-0.1775 - -0.0820
D11-Amp
0.3-30
9.1
0.0421
-0.0199
-0.0423 - +0.0157
D14-Met
0.3-30
20.5
0.0503
0.1226
+0.0637 - +0.1883
NMPhen
0.3-30
10.2
0.0449
+0.0260
-0.0075 - +0.0769
MDEA
N-PAmp
0.3-29
10.3
0.0264
-0.0597
-0.0879 - -0.0095
MDMA
D11-Amp
0.1-27
16.2
0.0503
-0.0750
-0.1423 - -0.0292
D14-Met
0.1-0.9(5)
15.4
0.0503(6)
-0.0712
-0.1247 - +0.0032
NMPhen
0.1-27
15.4
0.0496
-0.0722
-0.1136- -0.0108
0.1-10(5)
16.5
0.0379
-0.1030
-0.1414 - -0.0660
D14-Met
0.1-30
9.2
0.0351
-0.0343
-0.0767 - +0.0006
NMPhen
0.1-30
13.6
0.0322
-0.0827
-0.1221 - -0.0403
D11-Amp
0.1-10
17.7
0.0428
-0.1068
-0.1303 - -0.0586
D14-Met
0.1-3
11.3
0.0450
-0.0393
-0.0683 - -0.0205
NMPhen
0.1-3(5)
16.1
0.0449
-0.0871
-0.1279 - -0.0383
D11-Amp
0.1-30
12.8
0.0394
-0.0637
-0.0982 - -0.0433
D14-Met
0.1-30
9.8
0.0495
-0.0051
-0.0375 - +0.0556
NMPhen
0.1-30
11.0
0.0394
-0.0451
-0.0766 - -0.0163
D11-Amp
0.3-30
17.3
0.0402
-0.1073
-0.1496 - -0.0514
D14-Met
0.3-30
11.7
0.0519
-0.0294
-0.0559 - +0.0532
NMPhen
0.3-30
17.0
0.0450
-0.0956
-0.1197 - -0.0576
Caffeine
(L)-Ephedrine
(D)-Methamphetamine D11-Amp
Phencyclidine
Phentermine
Pseudoephedrine
(5)
(1) Backup Data Report [1]. Values are for chlorodifluoroacetyl derivatives and analysis by GC-MS in SIM mode (see Table 6 for MS conditions). Each sample consisted of a pair of 3” × 3” (7.5 cm x 7.5 cm) 12-ply cotton gauze pads. There were 6 replicate samples per concentration level and six concentration levels evaluated from approximately 0.1 to 30 μg/sample. Norephedrine (phenylpropanolamine) was not evaluated in the SIM mode due to breakdown at room temperature storage for several days prior to analysis. (2) Internal Standards, Deuterated: Non-deuterated: D11-Amp = Amphetamine-D11 NMPhen = N-Methyl phenethylamine
D14-Met = Methamphetamine-D14 N-PAmp = N-Propyl amphetamine
(3) Range used for calculation of precision, accuracy, and bias. The entire range studied for all analytes was approximately 0.1 to 30 μg/sample (1×LOQ to 300×LOQ). (4) NIOSH [1995]. NIOSH Technical Report: Guidelines for Air Sampling and Analytical Method Development and Evaluation. By Kennedy ER, Fischbach TJ, Song R, Eller PM, Shulman SA. Cincinnati, OH: U.S. Department of Health and Human Services, Public Health Services, Centers for Disease Control and Prevention, National Institute for Occupational Safety and Health, DHHS (NIOSH) Publication No. 95-117. (5) One or more higher level concentrations were omitted from the computations due to inlier CVs (<0.0200). (6) The overall precision, ŜrT, is an estimate due to inlier precisions (<0.02) at several higher concentration levels.
Method rev. 1.1.1
NIOSH Manual of Analytical Methods (NMAM), Fifth Edition